Subsections
9 Inductance
In addition to reading this assignment, read Appendix
C: Oscilloscope Guide. You may also need to refer to
Appendix A on uncertainties.
An inductor is a coil of wire which stores energy in a
magnetic field when it carries a current. The inductance
of an inductor is defined as the magnetic flux through the inductor
per unit current
 |
(25) |
Any change in the current through an inductor leads
to a change in the magnetic field it produces. This in turn leads to a
change in the magnetic flux through the inductor which, by Faraday's
law, produces an induced emf in the inductor.
Figure 20:
(a) A series LR circuit. (b) The potential difference across
the inductor after the switch is closed.
|
|
The moment the switch in the circuit shown in Figure 20
is closed, there is no current flowing, and the voltage across the
inductor is the same as the emf
of the voltage
source. The current immediately starts to increase from zero, changing
rapidly at first, and leveling off as it reaches its steady-state
value of
. The potential difference across the inductor
decays exponentially with the rate of change of the current as
 |
(26) |
where
is the time constant of the LR circuit given
by
 |
(27) |
This is the time it takes
to fall to a fraction
of its
initial charge. As you discovered in measuring capacitance in
Lab 7, it is easier in practice to measure the half-time
of an exponential decay than its time constant. The time
constant
is related to the half time
via
 |
(28) |
Just as the capacitance of a capacitor depends on the geometry of its
conductors, the inductance of an inductor depends on the geometry of
its wire loops. The inductance of a solenoid of cross sectional area
, length
, and having
turns is given by
 |
(29) |
You will use an oscilloscope to measure the inductance of the solenoid
you worked with in Lab 8. You will also investigate
the behavior of an LR filter. First, you will have a chance to
familiarize yourself with many of the controls and functions of an
oscilloscope (see Appendix C).
- Connect the alligator leads from channel 1 of the oscilloscope
across the output of the function generator.
- Set the function generator to produce a sinusoidal wave form.
- Set the triggering source to CH1.
- Trigger on a positive slope.
- Adjust the horizontal and vertical scale and position knobs and
triggering level knob so that the signal is stable, nice and big,
and centered in the display. If you run into trouble, ask for help.
- Experiment with the triggering level and slope controls until
you understand what they do. Under what circumstances does
triggering fail, giving a display like Figure 32 of
Appendix C?
- Disconnect the alligator leads from the function generator and
have someone hold one lead in each hand. Measure the frequency of
the signal. What could this be?
Figure 21:
(a) A circuit in which a function generator with internal
resistance
and a square wave form drives current back and forth
through an inductor with internal resistance
in series with a
resistor
.
(b) A qualitative plot of the potential difference across the
inductor vs. time.
|
|
- Measure and record the resistances
of the solenoid and
of the resistor with the DMM.
- You should find the internal resistance
of the function
generator on a label next to its output terminals. Record
them. (This can't be measured directly with the DMM.)
- Construct the circuit shown in
Figure 21. It is essential to connect the
black leads from the function generator and oscilloscope to the
opposite end of the solenoid from the resistor
, because they are
grounded.
- Connect the oscilloscope in parallel with the solenoid.
- Set the function generator to produce a square wave form.
- Adjust the vertical and horizontal scales and positions of the
oscilloscope to display the voltage across the inductor.
- Adjust the frequency of the function generator to obtain a
signal like that shown in Figure 21.
Note that the potential difference across the inductor
decays to a small constant value instead of zero. This is the
voltage drop across the internal resistance
of the solenoid.
- Measure the half-time
of the decay of the induced emf
in the solenoid. To get the best precision possible, adjust the
horizontal and vertical scales so that a single decay fills the
oscilloscope display.
One strategy:
- Use the vertical position knob to place the baseline
of the decay (the lower dashed line in
Figure 21) at the bottom of the display.
- Use a combination of the outer VOLTS/DIV knob and
inner variable scale knob to stretch the signal vertically so that
the maximum voltage (the upper dashed line in
Figure 21) is at the top of the display so that
the total drop in potential
across the inductor spans
the vertical scale of the display.
- The half-time of the decay is the time interval between the
peak and the point at which the signal crosses the center of the
display.
- For best precision, use the TIME/DIV knob to stretch
the signal as much as you can before making the half-time
measurement.
- Determine the inductance of the coil and its uncertainty from
your half-time measurement. Remember to use the sum of the
resistances of the resistor and the internal resistances of the
function generator and solenoid in your calculations!
- Use Eq. 29 and your measured inductance to
determine the number of turns
in the solenoid and the associated
uncertainty.
Discuss with your instructor preliminary answers to the questions
below.
... answers to the following.
- Under what circumstances could you get the display to look like
the mess in Figure 32 of Appendix C?
Explain why triggering fails.
- What do you think is the source of the signal you measured
across a person's hands? Explain.
- Give your results for the inductance and the number of turns of
the solenoid with uncertainties. Show the calculations you did to
get them.
|
Copyright © 2006-2009, L.A. Riley, T. J. Carroll, J.S. Scott
|
Updated Sun Apr 26 23:00:14 2009
|

This work is licensed under a Creative Commons License.